Xtra-Shield™ HS-714

HS-714

Xtra-Shield™ HS-714 is the preferred choice of leading medical device manufacturers worldwide for over 20 years for insulating electro-surgical instruments and for biocompatible jacketing for temporary implantation devices.

Request More Information

"*" indicates required fields

This field is hidden when viewing the form

Features and Benefits

* Including chlorinated cleaners, lubricating grease, penetrating oils, electrical insulation oils, and others.

Technical Specifications

  • Shrink Ratio: 2:1 NOTE: HS-714 can be customized to your specifications for expanded and recovered diameters, wall thicknesses and shrink ratios
  • Minimum Recommended Shrink Temperature: 121°C (250°F)
  • Maximum Recommended Shrink Temperature: 149°C (300°F)
  • Operating Temperature Range: -55°C to 121°C (-67°F to 250°F)
  • Longitudinal Shrinkage: +5/-5%
  • Standard Colors: Black, White, Blue, Clear

Physical / Electrical / Chemical Properties

  • Specific Gravity: 0.934
  • Durometer: Shore D 53
  • Tensile Strength: 3200 / 3600 psi
  • Elongation: 500 / 600%
    Secant Modulus: (2%) 1.2 X 104 psi
  • Heat Shock: 168 hours @ 150° C No dripping or cracking
  • Volume Resistivity: @ 500 volts >1016
  • Dielectric Strength: 1,800 / 2,000 vpm
  • Dielectric Constant @ 1MHz: 2.0 / 2.2
  • Dissipation Factor @ 1MHz: 0.008

Chemical Resistance

  • Resists a wide range of chemicals including

    • Hexachlorophene (pHisoHex)
    • saline solution
    • povidone-iodine (Betadine)
    • benzalkonium (Zephiran)
    • alcohol

    Resists moisture, UV light and fungus

AUTOCLAVE CYCLE TEST:
20 cycles @ 121°C 15 psig for 15 min. with fast dry cycle, followed by test for:
PropertyTest MethodTypical Value
Volume Resistivity @ 500 voltsASTM D 257No change from original
Dielectric Strength in air @ 500 kv/min riseASTM D 149No change from original
Dielectric Constant @ 1 MHzASTM D 150No change from original
Dissipation Factor @ 1 MHzASTM D 150No change from original

Notice to User: It is imperative that manufacturers and end-users understand that the electro-surgical instruments should be placed separately, without overlay, on autoclave trays. Potential damage caused by overlay in the autoclave will significantly reduce this and any other dielectric insulator’s ability to pass the 20 cycles of autoclave required by the AAMI.

All statements and recommendations are based on technical data that INSULTAB believes to be true and reliable. It is the responsibility of the user to determine functionality and suitability. If you have any questions on the use or handling of this product, please contact us at 800-468-4822.

Standard Dimensions

Expanded I.D. Minimum Recovered I.D. Maximum Recovered Wall Nominal Standard Packaging (per Box) Product is Supplied in Coils or on Plastic Spools
Size in. mm in. mm in. mm Ft. / Coil or Spool Total Ft. / Box
3/32 .093 2.36 .046 1.17 .020 0.51 1,000′ 2,000′
1/8 .125 3.18 .062 1.58 .020 0.51 1,000′ 2,000′
3/16 .187 4.75 .093 2.36 .020 0.51 1,000′ 2,000′
1/4 .250 6.35 .125 3.18 .025 0.64 500′ 1,000′
3/8 .375 9.53 .187 4.75 .025 0.64 200′ 600′
1/2 .500 12.70 .250 6.35 .025 0.64 200′ 600′

We will manufacture to your precise specifications

All information presented is believed to be reliable and is offered only as a guide to product selection. As each application is unique, Insultab can make no warranties as to the suitability of any products for a particular use. Specifications are subject to change.

 