Insultab 714 // Medical Grade Acrylated Olefin Tubing

The preferred choice of leading medical device manufacturers worldwide for over 20 years for insulating electro-surgical instruments and for biocompatible jacketing for temporary implantation devices.

Uses and Features

  • Manufactured using a proprietary, USP Class VI certified resin with no additives or fillers
  • Offers a bio-compatible jacketing for temporary implantation devices , Ideal for insulating electro-surgical instruments
  • Offers unique adhesion properties to metals, glass, and select plastics without the use of adhesive
  • Displays superior cut-though, puncture, and abrasion resistance
  • Provides excellent resistance to a wide range of chemicals including Saline, Povidone-iodine (Betadine), Benzalkonium (Zephiran), and Alcohol
  • Can be printed with USP Class VI inks
  • Comes with complete lot traceability
  • Alcohol wiped and double-bagged in a clean room environment

Technical Data

  • Shrink ratio: 2:1
  • Recommended Shrink Temperature: 121°C
  • Operating Temperature Range: -20°C to 121°C
  • Longitudinal Change: + / -5%
  • Shelf Life: 3 years
  • Recommend Storage: <27⁰C
  • Standard Colors: Clear, White, Black, Blue
  • Standard Packaging: Spooled

Certifications

  • USP Class VI
  • 1SO-10993-5
  • USFDA Reg C.F.R 177.1330
  • RoHS Compliant

USP Class VI and ISO 10993-5 Cytotoxicity documentation
provided as post process certification when requested

CategoryPropertyUnitsValue
MechanicalSpecific Gravityg/cm20.934
Tensile Strength (min)psi3200
Ultimate Elongation (min)%500
DurometerShore D53
ElectricalDielectric Strength (min)v/mil1800
Flammabilityv/milVW-1

E-beam and GammaSteris Process — Peracetic Acid
AutoclaveHydrogen Peroxide Gas
EtOGlutaraldehyde

20 cycles @ 121°C 15 psig for 15 min. with fast dry cycle, followed by test for:

PropertyTest MethodTypical Value
Volume Resistivity @ 500 voltsASTM D 257No change from original
Dielectric Strength in air @ 500 kv/min riseASTM D 149No change from original
Dielectric Constant @ 1 MHzASTM D 150No change from original
Dissipation Factor @ 1 MHzASTM D 150No change from original

Notice to User: It is imperative that manufacturers and end-users understand that the electro-surgical instruments should be placed separately, without overlay, on autoclave trays. Potential damage caused by overlay in the autoclave will significantly reduce this and any other dielectric insulator’s ability to pass the 20 cycles of autoclave required by the AAMI.

All statements and recommendations are based on technical data that PEXCO believes to be true and reliable. It is the responsibility of the user
to determine functionality and suitability. If you have any questions on the use or handling of this product, please contact us at 844-352-5777.

SizeExpanded I.D. MinimumRecovered I.D. MaximumRecovered Wall NominalStandard Packaging (per Box) Product is Supplied in Coils or on Plastic Spools
in.mmin.mmin.mmft. / Coil or SpoolTotal ft./Box
3/32.0932.36.0461.17.0200.511,000′2,000′
1/8.1253.18.0621.58.0200.511,000′2,000′
3/16.1874.75.0932.36.0200.511,000′2,000′
1/4.2506.35.1253.18.0250.64500′1,000′
3/8.3759.53.1874.75.0250.64200′600′
1/2.50012.70.2506.35.0250.64200′600′

Custom sizes available upon request

All information presented is believed to be reliable and is offered only as a guide to product selection. As each application is unique, Insultab can make no warranties as to the suitability of any products for a particular use. Specifications are subject to change.

Insultab 714 // Medical Grade Acrylated Olefin Tubing